Thermoworks 9009-B INDUSTRIAL DUAL-BLOCK CALIBRATOR Instrukcja Użytkownika

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Urządzenia pomiarowe Thermoworks 9009-B INDUSTRIAL DUAL-BLOCK CALIBRATOR. ThermoWorks 9009-B INDUSTRIAL DUAL-BLOCK CALIBRATOR User Manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj

Podsumowanie treści

Strona 1 - User’ s Guide

9009Industrial Dual-Block CalibratorUser’ s GuideRev. 6A0901Hart Scientific

Strona 2 - Rev. 6A0901

If this equipment is used in a manner not specified by the manufacturer, theprotection provided by the equipment may be impaired or safety hazards may

Strona 3 - Table of Contents

ELECTRICAL HAZARDThese guidelines must be followed to ensure that the safety mechanisms in thisinstrument will operate properly. This instrument must

Strona 4

DO NOT operate this instrument in an excessively wet, oily, dusty, or dirtyenvironment.The unit is a precision instrument. Although it has been design

Strona 5

Phone: +31-402-675300Telefax: +31-402-675321E-mail: [email protected] Int'l CorporationService Center - InstrimpexRoom 2301 Sciteck Tower

Strona 6

2 IntroductionThe Hart Scientific 9009 Industrial Dual-Block Calibrator may be used as aportable instrument or bench top temperature calibrator for ca

Strona 7

3 Specifications and EnvironmentalConditions3.1 SpecificationsSpecification Hot Block Cold BlockRange50°C to 350°C (122°F to 662°F) –15°C to 110°C (5°

Strona 8

•ambient relative humidity: maximum 80% for temperature <31°C, de-creasing linearly to 50% at 40°C•pressure: 75kPa - 106kPa•mains voltage within ±

Strona 9 - 1.2 Safety Information

4 Quick Start4.1 UnpackingUnpack the calibrator carefully and inspect it for any damage that may have oc-curred during shipment. If there is shipping

Strona 10 - Safety Information

requirements. Turn the calibrator on using the “POWER” switch on the powerentry module (PEM). The calibrator will turn on and begin to heat or cool to

Strona 11 - CAUTIONS

5 Parts and ControlsThe user should become familiar with the dry-well calibrator and its parts.5.1 Top Panel (Lid Open)Power Switch - The power switch

Strona 12 - Authorized Service Centers

Limited Warranty & Limitation of LiabilityEach product from Fluke’s Hart Scientific Division (“Hart”) is warranted to be free from defects in mate

Strona 13 - User’s Guide

Serial Port - A three-conductor jack is present for interfacing the calibrator toa computer or terminal with serial RS-232 communications.Controller D

Strona 14 - 2 Introduction

5.3 AccessoriesThe following inserts are available for the 9009 calibrator. Inserts may be usedwith either the hot or cold temperature wells.Model Des

Strona 15 - Conditions

6 General Operation6.1 Changing Display UnitsThe 9009 can display temperature in Celsius or Fahrenheit. The temperatureunits are shipped from the fact

Strona 16

7 Controller OperationThis chapter discusses in detail how to operate the calibrator temperature con-troller using the front control panel. Using the

Strona 17 - 4 Quick Start

9009 Industrial Dual-Block CalibratorUser’s Guide20Figure 3 Controller Operation Flowchart

Strona 18 - 4.5 Changing Display Units

100.0 CWell temperature in degrees CelsiusSAccess set-point memory1 100.Set-point memory 1, 100°C currently usedTo change to another set-point memory

Strona 19 - 5 Parts and Controls

Un= F New units selectedPress “SET” to accept the present setting and to continue.7.3 ScanThe scan rate can be set and enabled so that when the set-po

Strona 20

7.4 Secondary MenuFunctions which are used less often are accessed within the secondary menu.The secondary menu is accessed by pressing “SET” and “EXI

Strona 21 - 5.3 Accessories

set-point. This range of temperature is called the proportional band. At the bot-tom of the proportional band the heater output is 100%. At the top of

Strona 22 - 6 General Operation

SAccept the new proportional band setting7.7 Controller ConfigurationThe controller has a number of configuration and operating options and calibra-ti

Strona 23 - 7 Controller Operation

Table of Contents1 Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . 11.1 Symbols Used . . . . . . .

Strona 24

7.9.1 BAUD RateThe BAUD rate is the first parameter in the menu. The BAUD rate setting de-termines the serial communications transmission rate.bAUdFla

Strona 25 - Temperature Set-point

The mode may be changed using “UP” or “DOWN” and pressing “SET”.d=HALFNew duplex mode setting7.9.4 LinefeedThe final parameter in the serial interface

Strona 26 - 7.3 Scan

parameter to the new value. Pressing “EXIT” causes the parameter to beskipped ignoring any changes that may have been made.7.10.1 R0This probe paramet

Strona 27 - 7.6 Proportional Band

8 Digital Communication InterfaceThe calibrator is capable of communicating with and being controlled by otherequipment through the digital serial int

Strona 28

8.1.2 SetupBefore operation the serial interface must first be set up by programming theBAUD rate and other configuration parameters. These parameters

Strona 29 - 7.8 Operating Parameters

8.1.3 Serial OperationOnce the cable has been attached and the interface set up properly the control-ler immediately begins transmitting temperature r

Strona 30

9009 Industrial Dual-Block CalibratorUser’s Guide32Command DescriptionCommandFormat†CommandExample‡ReturnedReturnedExampleAcceptableValuesDisplay Temp

Strona 31 - 7.10 Calibration Parameters

338 Digital Communication InterfaceInterface CommandsCommand DescriptionCommandFormat†CommandExample‡ReturnedReturnedExampleAcceptableValuesCalibratio

Strona 32 - 7.10.4 BETA (Low Temp Only)

9 Test Probe CalibrationFor optimum accuracy and stability, allow the calibrator to warm up for 10minutes after power-up and then allow adequate stabi

Strona 33 - 8.1 Serial Communications

made at the desired temperatures with the desired test probes to establish thesetimes.9009 Industrial Dual-Block CalibratorUser’s Guide36

Strona 34 - 8.1.2 Setup

7.3.2 Scan Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 227.4 Secondary Menu. . .

Strona 35 - 8.2 Interface Commands

10 Calibration ProcedureSometimes the user may want to calibrate the instrument to improve the tem-perature set-point accuracy. Calibration is done by

Strona 36

CTTTT=⎡⎣⎢⎤⎦⎥−⎡⎣⎢⎤⎦⎥−⎡⎣⎢⎤⎦⎥−⎡⎣⎢⎤⎦44331001100 1001100⎥DTTTT=⎡⎣⎢⎤⎦⎥−⎡⎣⎢⎤⎦⎥−⎡⎣⎢⎤⎦⎥−⎡⎣⎢⎤⎦33221001100 1001100⎥ER R=−43FRR=−32deltaAF BEDE CF=−−Where:T1-4- M

Strona 37 - Interface Commands

yT=⎡⎣⎢⎤⎦⎥1100betaapha x yTxydeltayRrzeroalph=+−−131321( )()()()() ( ax y)( )( )3Where:t and r are the measured resistance at –15°C.alpha, rzero, and d

Strona 38 - 9 Test Probe Calibration

11 Maintenance•The calibration instrument has been designed with the utmost care. Easeof operation and simplicity of maintenance have been a central t

Strona 39

12 Troubleshooting12.1 TroubleshootingIn the event that the dry-well appears to function abnormally, this section mayhelp to find and solve the proble

Strona 40 - 10 Calibration Procedure

12.2.2 Low Voltage Directive (Safety)In order to comply with the European Low Voltage Directive (73/23/EEC),Hart Scientific equipment has been designe

Strona 41 - 10.2.2 Compute R0 & ALPHA

12.1 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . 4312.2 Comments . . . . . . . . . . .

Strona 42 - Calibration Procedure

ivFiguresFigure 1 Top Panel (Lid Open) . . . . . . . . . . . . . . . . . . . . . . . . . . 13Figure 2 Constant Temperature Bl

Strona 43 - 11 Maintenance

vTablesTable 1 International Electrical Symbols . . . . . . . . . . . . . . . . . . . . . 1Table 2 9009 Controller Communications

Strona 44 - 12 Troubleshooting

1 Before You Start1.1 Symbols UsedTable 1 lists the International Electrical Symbols. Some or all of these symbolsmay be used on the instrument or in

Strona 45

Symbol DescriptionCanadian Standards AssociationOVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 re-fers to the level of Impul

Komentarze do niniejszej Instrukcji

Brak uwag